Aggregate Analysis by SEC

Size-Exclusion HPLC for Peptide Aggregates

Principle

Size-exclusion chromatography (SEC) separates molecules by hydrodynamic volume. Larger aggregates elute first; monomers elute later. SEC is the primary method for aggregate quantification in peptide products.


Method Parameters

Parameter Recommended Value
Column Silica-based diol (200–300 Å pore)
Mobile phase 100 mM phosphate + 200 mM NaCl, pH 7.0
Flow rate 0.5 mL/min
Temperature 25 °C (ambient)
Detection UV 214, 280 nm
Injection 10–50 µg
Run time 30–40 min
Calibration Molecular weight standards

Aggregate Classification

Species Apparent MW Elution Order Concern Level
Soluble oligomers Dimer–hexamer Before monomer Medium
Sub-visible particles 0.1–25 µm Filtered High
Visible particles >25 µm Filtered Critical
Monomer Expected MW Main peak
Fragments <Expected MW After monomer Low

Acceptance Criteria

Application Monomer Aggregate Fragment
Research peptide ≥95% ≤3% ≤2%
GMP peptide (parenteral) ≥98% ≤2% ≤1%
Peptide therapeutic ≥99% ≤1% ≤0.5%