Skip to content

Charge Profile Analysis

Ion Exchange Chromatography for Peptide Charge Variants

Purpose

Charge profile analysis detects charge variants (deamidation, C-terminal truncation, succinimide formation) that may not be resolved by reversed-phase HPLC.


Method Comparison

Method Principle Detects
CEX (Cation Exchange) Positive charge binding Deamidation, succinimide
AEX (Anion Exchange) Negative charge binding C-terminal variants
IEF (Isoelectric Focusing) pI-based separation Complex charge mixtures
icIEF (Imaged cIEF) Real-time IEF High-resolution charge profile

CEX Method Parameters

Parameter Typical Value
Column Wheat germ agglutinin (WGA) or sulfopropyl
Mobile Phase A 20 mM MES, pH 5.5–6.5
Mobile Phase B A + 0.5 M NaCl
Gradient 0–50% B in 30 min
Detection UV 214, 280 nm
Flow rate 0.5–1.0 mL/min
Column temp 30 °C

Charge Variants in Peptides

Variant Charge Change CEX Retention Cause
Deamidation (Asn→Asp) −1 Earlier pH/temperature stress
Succinimide −1 Earlier Asp/Gly sequences
C-terminal truncation Varies Varies Process impurity
Oxidation 0 (no change) Slight Met/Trp exposure
Aggregation Varies May not elute Concentration